Nacalai USA - Innovations for Life Sciences

Subtotal: $0.00

View Cart

COSMOSIL Buckyprep-M

Specialty for Fullerenes / For Metallofullerenes

COSMOSIL Buckyprep-M is a phenothiazinyl group bonded silica based column specifically designed for metallo-fullerene separation. Metallo-fullerenes are retained more strongly than other fullerenes on this column. COSMOSIL Buckyprep-M is also effective for the separation of higher fullerenes and fullerene derivatives.

COSMOSIL Buckyprep - M
Silica gel High purity porous spherical silica gel
Average particle size 5 µm
Average pore size approx. 120Å
Specific surface area approx. 300m²/g
Stationary phase Phenothiazinyl
End capping treatment No
Carbon content approx. 13%

 

Application 1

Column size: 4.6mmI.D.×250mm; Mobile phase: toluene; Flow rate: 1.0 ml/min.;
Detection: UV 312 nm; Temperature: 30°C
Data courtesy of Professor Hisanori Shinohara (Department of Chemistry, Nagoya University)

Application 2

Column size: 4.6mmI.D.×250mm; Mobile phase: toluene; Flow rate: 1.0 ml/min.;
Detection: UV 312 nm; Temperature: 30°C
Data courtesy of Professor Hisanori Shinohara (Department of Chemistry, Nagoya University)

Usage Information

Column5 µm Buckprep-M
I.D. (mm)2.04.610.020.0
Washing method Wash with 1,2,4,-trichlorobenzene or other strong solvent to remove adsorbed contaminants.
Storage conditions Wash column, then replace solvent with toluene. Plug tightly and store in a cool, dry place.
Max. pressure 20MPa 15MPa
Temperature range The maximum usable temperature is 60°C. However. for regular use, please use at a constant temperature between 20°C and 50°C.

 

 

 

 

 

 

 

 

Solvents usable with Buckyprep Series
SolventFeaturesC60 Solubility (mg/ml)
Toluene General-use solvent for fullerenes. 3.2
n-Hexane Weaker eluent than toluene. 0.046
n-Heptane **
Methanol 0.001
2-Propanol **
Acetonitrile Weaker eluent than toluene. 0.018
Chlorobenzene Stronger eluent than toluene. 7
o-Dichlorobenzene Stronger eluent than chlorobenzene. 27
1,2,4-Trichlorobenzene Strongest commonly-used eluent. Can be used to flush higher fullerenes. 21.3

Fullerene Chromatogram Index

Fullerene Chromatogram Index includes more than 100 chromatograms is available at the website of The Fullerenes, Nanotubes and Graphene Research Society. (Click here to download)

FullereneChromatogramIndex

COSMOSIL Application Search

COSMOSIL Application, which includes more than 7,700 applications using COSMOSIL/COSMOCORE columns, is now usable by clicking the link below which is in Nacalai Tesque, Inc. The application is searchable by sample category, sample name, CAS No., column name and particle size

Ordering Information

Product Cat.No. End-Fitting PKG Size Price(US$)  
COSMOSIL Buckyprep-M Packed Column (4.6 mm I.D. x 250 mm) 04138-71 Waters type 1 pkg 1,980.00 Buy
COSMOSIL Buckyprep-M Packed Column (10.0 mm I.D. x 250 mm) 04141-11 Waters type 1 pkg 6,447.00 Buy
COSMOSIL Buckyprep-M Packed Column (20.0 mm I.D. x 250 mm) 04142-01 Waters type 1 pkg 24,043.00 Buy
COSMOSIL Buckyprep-M Packed Column (28.0 mm I.D. x 250 mm) 05873-01 Waters type 1 pkg 36,584.00 Buy
COSMOSIL Buckyprep-M Guard Column (4.6 mm I.D. x 10 mm) 04139-61 Waters type 1 pkg 519.00 Buy
COSMOSIL Buckyprep-M Guard Column (10.0 mm I.D. x 20 mm) 04140-21 Waters type 1 pkg 1,381.00 Buy
COSMOSIL Buckyprep-M Guard Column (20.0 mm I.D. x 50 mm) 34474-31 Waters type 1 pkg 6,014.00 Buy
COSMOSIL Buckyprep-M Guard Column (28.0 mm I.D. x 50 mm) 05872-11 Waters type 1 pkg 11,573.00 Buy